The CALYPSO DS.HP Nitrogen generator has been specifically designed to meet the gas flow, purity and pressure requirements of ASE instruments.The Nitrogen generator uses pressure swing adsorptiontechnology (PSA) combined with a pressure management in two steps to produce pure nitrogen gas with a high flow, high pressure and purity.The CALYPSO DS.HP is available with integrated oil free air compressor using the Energy Saving Technology mode (EST). It is extremely compact and quiet in operation.The generator is controlled using the latest in color touch screen technology to display the process in real time, inlet / outlet pressure, N2 Flow and status of the system with remote access for diagnostic control.
CALYPSO DS.HP Nitrogen Generator
Double Step PSA High Pressure Technology
Reference | CALYPSO.DS.HP.[Flow rate]/1 (With compressor)
Nitrogen
ASE-Dionex
Advantages product
IMPROVE ANALYTICAL INSTRUMENTS PERFORMANCE
Improve analytical instruments performance
IMPROVE LABORATORY EFFICIENCY
The relatively high gas volumes required by LCMS instruments make cylinder supply inappropriate for such applications. A constant, uninterrupted gas supply eliminates interruptions of analyses to change cylinders.
IMPROVE ECONOMY
Quick return on investment < 1 year. No gas cylinder rental bottles, no price inflation. The 2-stage pressure design allows the two compressors to work at their optimum pressure range which is reducing the stress on the compressors and as result extending the life time of the compressors - which is a considerable saving.
IMPROVE SAFETY
Nitrogen produced at low pressure and ambient temperatureremoves the hazards associated with high pressure cylindersand liquid Dewarโs.
Characteristics
Intelligent color touch screen to provide simple and user friendly management of all functions of the unit
Fully regenerative PSA technology in Two Step (DS-PSA)
High Nitrogen pressure up to 11 bar (159 psi)
With integral oil free air compressor
Quiet thanks to the Soundproofed compressor box and anti-vibration features
Auto start
Audible alarm display and Visual maintenance indication with history log
Outlet flow indicator
Energy Saving Technology (EST) :
- Switch off the compressor when gas supply is not required
- Allowing low energy consumption and increasing durability of the compressor
- Switch off the compressor when gas supply is not required
- Allowing low energy consumption and increasing durability of the compressor
Remote PC monitoring and diagnostic analysis via USB to interface the unit with customerโs PC software (allow to carry out checks and maintenance effectively, only via a remote connection)
Compact design
Fit with wheels
Specifications
Models | CALYPSO.DS.HP.40/1 | |||
---|---|---|---|---|
GENERALS INFORMATIONS | ||||
Max Nitrogen Flow rate | 40 L/min | |||
Max Nitrogen outlet pressure | 11 bar (159 psi) | |||
N2 purity | > 99% | |||
Particles | 0.01 ฮผ | |||
Touch screen | Touch screen (operating parameters, system status, alarms) with LED indicators(Power on/off; ready or errors) | |||
Max Relative Humidity | 80% RH @ 25ยฐC (77ยฐF) | |||
Temperature range | From 5 – 35ยฐC (41 – 95ยฐF) | |||
Noise level @ 1M | < 55 dB (A) | |||
Dimensions(W x H x D) | 47.5 x 66.2 x 79 cm (18.7″ x 26″ x 31″) | |||
Weight (kg/lbs) | 125 / 275 | |||
Electrical supply | 220 – 240 V ac / 1 ph / 50 – 60 Hz | |||
110 – 120 V ac / 1 ph / 50 – 60 Hz | ||||
Power consumption | 1820 W | |||
CONNECTIONS | ||||
Nitrogen outlet | 1/4″ GF | |||
COMMUNICATION | ||||
USB/PC Control | In series |