STREAM 3G

N2/Dry Air Generator for Sciex LC-MS

Azote/Air sec
Sciex LC-MS

 

The Membrane Nitrogen/Air generator has been developed to meet specific requirements in terms of flow, purity and pressure for Sciex LC-MS.

Delivering one output of laboratory-grade nitrogen and two dry air outputs, the generators are available with different gas flows.

F-DGS’s STREAM 3G series generators have been engineered to provide Nitrogen and air to laboratories which already have an external source of clean, dry air.

These systems contain no moving parts, are effectively silent whilst in operation and require minimal maintenance

Reference :

STREAM.[Flow rate].3G

Avantages

 

IMPROVE LABORATORY EFFICIENCY

The relatively high gas volumes required by LCMS instruments make cylinder supply inappropriate for such applications. A constant, uninterrupted gas supply eliminates interruptions of analyses to change cylinders.

 

IMPROVE ANALYTICAL INSTRUMENTS PERFORMANCE

Production of a constant flow of gas improves the consistency of the analysis results and hence reproducibility.

 

IMPROVE ECONOMY

– Quick return on investment < 1 year
– No gas cylinder rental bottles, no price inflation

 

IMPROVE SAFETY

Nitrogen and dry air produced at low pressure and ambient temperature removes the hazards associated with high pressure cylinders and liquid Dewar’s

 

SIMPLE INSTALLATION

Gas generators can be installed in the laboratory, on the wall, eliminating the need for long gas lines from cylinders secured elsewhere. No power supply is require.

Caractéristiques

  • Capable of supplying multiple Sciex LC-MS instruments
  • 24h/7 operation at optimum performance if required
  • Gas is supplied on demand
  • Completely silent in operation
  • Wall mounted installation : Save space in the lab
  • No noise : No mechanical parts moving
  • No need of electricity supply
  • Low maintenance : only to replace the filters once per year

Spécifications

MODELS STREAM.12.3G STREAM.27.3G STREAM.54.3G STREAM.85.3G
GENERALS INFORMATIONS
Max N2 Flow rate 12 l/min @ 4.5 bar
(65 psi)
27 l/min @ 4.5 bar
(65 psi)
54 l/min @ 4.5 bar
(65 psi)
85 l/min @ 4.5 bar
(65 psi)
Dry air outlet flow for the source gas 26 l/min @ 7.6 bar
(110 psi)
46 l/min @ 7.6 bar
(110 psi)
55 l/min @ 7.6 bar
(110 psi)
85 l/min @ 7.6 bar
(110 psi)
Dry air outlet flow for the exhaust gas 16 l/min @ 4.1 bar
(60 psi)
23 l/min @ 4.1 bar
(60 psi)
46 l/min @ 4.1 bar
(60 psi)
60 l/min @ 4.1 bar
(60 psi)
Air inlet pressure Min./Max. 8 – 10 bar (116 – 145 psi)
Pressure loss < 0.8 bar (10 psi)
N2 dewpoint at operating pressure -40°C (-40°F)
Particles < 0.01 ppm
Air flow rate required @ 8 bar (116 psi) 115 l/min 240 l/min 330 l/min
Suspended liquids None
Temperature range 5 – 30°C (50 – 95°F)
Dimensions (W x H x D) 38 x 120 x 22.5 cm (15″ x 47″ x 9″)
Weight (kg/lbs) 24 / 53 30 / 66
Electrical supply 220 – 240 V ac / 1 ph / 50 – 60 Hz
110 – 120 V ac / 1 ph / 50 – 60 Hz