RACK.MB.H2
Générateur d'Hydrogène 19"
The RACK-H2 hydrogen generators use the exclusive electrolyte membrane technology (PEM cell) 100% full titanium to produce high grade purity H2 gas.
The exclusive cold dual dynamic regeneration dryer on RACK.MF series is completely maintenance free and allows continuous operation 24 hours a day. The automatic checking for internal leaks whenever starting the unit and constant control of operating paameters guarantee maximum safety.
The drying system on the RACK.MB.H2 Series use a Drying system : static, maintenances-free
Reference : RACK.MB.H2 (H2 flow rate)

Avantages
Improved chromatograph result
The use of hydrogen as a carrier gas allows lower temperature elution, thus extending the life of the chromatograph column. Hydrogen as a carrier gas is faster and more sensitive than the
more-expensive helium. Run time savings of 25% to 35% without a decline in resolution.
Increased laboratory efficiency
A constant, uninterrupted gas supply of guaranteed purity eliminates interruptions of analyses to change cylinders and reduces the amount of instrument re-calibrations required.
Improved safety
The very limited internal volume (less than 50 ml) allows safe use of the gas generators where the use of cylinders is risky or prohibited. The application of tested safety technologies stops the unit in the event of leaks or malfunctions
Simple installation
Gas generators can be installed in the laboratory, on or under a bench, eliminating the need for long gas lines from cylinders secured elsewhere.
Caractéristiques
- RACK.MB.H2: use a Drying system: static, maintenances-free
- Flow rate: 100 to 1400 cc/min
- Pressure up to 11 bar (160 psi)
- LCD touch screen: with indication in real time: H2 outlet pressure, H2 flow rate, water quality, water level, system status with auto-diagnostics of breakdown with alarms
- Remote PC monitoring via RS232/RS485/USB to interface the unit with customer's PC software (allow the end user to be connected with the manufacturer for remote diagnostics)
- Up to 32 units can be controlled in cascading mode
Spécifications
MODELS | RACK.MB.H2.100 | RACK.MB.H2.160 | RACK.MB.H2.250 | RACK.MB.H2.300 | RACK.MB.H2.500 | RACK.MB.H2.600 | |||
---|---|---|---|---|---|---|---|---|---|
GENERALS INFORMATIONS | |||||||||
H2 flow rate – cc/min | 100 | 160 | 250 | 300 | 500 | 600 | |||
Electrolytic cell | Long life Multilayer PEM technology | ||||||||
H2 purity | > 99.99999% (O2 < 0.1 ppm, Dew Point H2O < -75°C (-103°F)) | ||||||||
Delivery pressure | 0.1 – 11 bar (1 – 160 psi) | ||||||||
H2 dryer | Drying system : static, maintenances-free | ||||||||
Temperature range | From 5 – 35°C (41 – 95°F) and humidity 80% to 25°C (77°F) | ||||||||
LCD digital screen | LCD digital screen (operating parameters, system status, alarms) with LED indicators (Power on/off; ready or errors) | ||||||||
Dimensions (W x H x D) | Standard 19” Rack 3U – depth 54 cm | ||||||||
Weight (kg/lbs) | 17 / 37 | ||||||||
Electrical supply | 220 – 240 V ac / 1 ph / 50 – 60 Hz | ||||||||
110 – 120 V ac / 1 ph / 50 – 60 Hz | |||||||||
WATER | |||||||||
Water quality | Deionised, ASTM II, < 0.1 µS Conductivity | ||||||||
Internal water tank | 300 ml with automatic refill via an internal sucking water pump | ||||||||
External water tank capacity (Optionnal) | 5 or 7 liters | ||||||||
CONNECTIONS | |||||||||
Hydrogen outlet | 1/8 Swagelock | ||||||||
Water inlet | Quick release push-in fitting | ||||||||
COMMUNICATION | |||||||||
USB/PC Control | In series | ||||||||
RS232/RS485 | Option | ||||||||
SPECIAL FUNCTION | |||||||||
Cascading mode | Option | ||||||||
Automatic water tank refill | Option |
ACCESSORIES | Water tank 5 liters | Water tank 10 liters |
---|---|---|
Part Number | water.tank.005 | water.tank.006 |